Kippzonen CGR 4 Pyrgeometers Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 35
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
For the CGR 4 the effect of each parameter on the sensitivity can be shown separately. The non-linearity
error, the sensitivity variation with irradiance, is the same for any CGR 4 and is shown in Figure 8.
Figure 8: Non-linear sensitivity variation with irradiance of a CGR 4 pyrgeometer
The temperature dependence of the sensitivity is a function of the individual CGR 4. For a given CGR 4 the
response lies in the region between the curved lines in Figure 9. The temperature dependence of each CGR
4 pyrgeometer is characterised and supplied with the instrument. Each CGR 4 has a built-in temperature
sensor to allow corrections to be applied if required.
Figure 9: Typical temperature dependency of a CGR 4
Page 14 CGR 4 Manual
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 34 35

Comentarios a estos manuales

Sin comentarios