17
2 LIST OF SPECIFICATIONS
Spectral range: 300 to 2800 nm, 50% points
Sensitivity: 7 µV/Wm
-2
(nominal)
Impedance: 500 to 2000 Ω
Response time: 18 s (95% response)
< 8 s (63% response)
Non-linearity: Max. 3 % (0 - 2500 W/m
2
)
Directional error (at 80°
with a 1000 W/m² beam): ± 20 W/m²
Temperature dependence
of sensitivity: 3 % (-20 °C to +0 °C)
2 % (0 °C to +100 °C)
3 % (+100 °C to +150 °C)
Tilt error: Max. 1% deviation when facing
downwards
Zero-offset due to
temperature changes: Max. 4 W/m
2
offset for 5 K/h temp.
change.
Zero-offset due to FIR
(200 W/m
2
): ± 15 W/m
2
Operating temperature: -40 °C to +150 °C
Field of view: 180° (2 π sr)
Max irradiance: 4000 W/m
2
Non-stability: ± 1% sensitivity change per year