Pyrheliometer Specification List
Specification Class of Pyrheliometer
Secondary
Standard
First Class Second Class
Response time: time for 95% response < 15 s < 20 s < 30 s
Zero off-set:
response to 5 K h change in ambient
-1
temperature
± 2 Wm ± 4 Wm ± 8 Wm
-2 -2 -2
Resolution (smallest detectable change in Wm ) ± 0.5 Wm ± 1 Wm ± 5 Wm
-2 -2 -2 -2
Stability (percentage of full scale, change/year) ± 0.5 % ± 1 % ± 2 %
Non-linearity:
percentage deviation from the responsivity
at 500 Wm due to change in irradiance within
-2
100 Wm to 1000 Wm
-2 -2
± 0.2 % ± 0.5 % ± 2 %
Spectral selectivity:
percentage deviation of the product of the
spectral absorptance and the spectral
transmittance from the corresponding mean
within 0.3 :m and 3.0 :m
± 0.5 % ± 1 % ± 5 %
Temperature response:
total percentage deviation due to change in
ambient temperature within in interval of 50 K
± 1 % ± 2 % ± 10 %
Tilt response:
percentage deviation form the responsivity at 0°
tilt (horizontal) due to change in tilt from 0° to
90°at 1000 Wm irradiance
-2
± 0.2 % ± 0.5 % ± 2 %
Traceability: maintained by periodic comparison with a primary
standard
pyrheliometer
with a
secondary
standard
pyrheliometer
with a first
class
pyrheliometer
or better
Table B 1.2. Pyrheliometer specification table from ISO 9060.