111
Cavity size; CWL tolerance 3-cavity, Ø25 mm; ±2 nm
Side-band blocking OD4, UV to 1200 nm
Detector type; active area UV si-photodiode; 33 mm2
Sensitivity gain setting x 4 channels high/low by jumper; trim via multi-turn pot
Output signal x 4 channels -0.05 to 4.5 VDC m ax.
Resolution <0.005OD (Langley method)
Response time 0.2 s to 99%
Operating temperature -30 °C to +70 °C
Power supply requirement 5.5 to 14.5 VDC, 20 mA
Temperature output 10 mV °C (0.20 V = 20 °C)
-1
Desiccant silica gel (visible through front window)
Interface lead 8-core shielded cable, 5 m (free end is unterminated)
Mounting method Ø25 mm (1") female clamp, with adjustable alignment
Weight 1.25 kg (excluding 0.5 kg lead)
B 5.7 Carter-Scott Design Middleton SP01 Sunphotometer
The SP01 version has fixed field stops and operates in sun photom eter mode only.
• Four narrow-bandwidths measured simultaneously
• Detectors and filters held at constant temperature to enhance stability
• Low- noise signal amplifiers with negligible zero offset drift
• Three gain selections accommodate a wide range of signal strengths
• Motorised field stop on SP01-A can pause for signal zero check
• Weather-proof aluminum housing
Specifications
(A) Spectro-Pyrheliometer Tube
Dimension 48 x 312 nm
Field of view: sun photometer field stop 2.4° (1.2° opening)
solar aureole field stop opening angle 3° - 5°, centre-blocked
(SP01-A only)
Slope angle (with 2.4° field stop) 0.9° (limit angle = 1.5°)
Filter wavelength and bandpass 368BP5nm; 412BP5; 500BP5; 862BP5
(Optional filters) 675BP5; 778BP5; 812BP5
Filter size and CWL tolerance Ø25 mm, ± 2 nm
Sensitivity gain selection (by jumper) low (= medium x 0.2); medium; high (= medium x 5)
Output signal -0.05 to +8.0 VDC max
Response time 2 s to 99%