50
- for friction-driven drives check for slippage of the drive disks (see tracker operating
manual for the proper procedure).
- if slippage occurs on gear driven trackers, the gears should be inspected for missing
teeth and the gear alignment tested (see tracker operating manual for proper procedures).
- check to ensure that the tracker has not changed its physical position, either in level
or location (e.g., the tracker has not been bumped accidentally).
- a tracker mechanical malfunction or software failure can also cause a loss of tracking
capability. The operator should refer to the tracker operating manual in such cases.
The operator should always note the position of the solar spot on the pyrheliometer or cavity
radiometer before any adjustment is made. Following the adjustment, the new location of
the solar spot should be noted. The time required to make the adjustment and the details
of what caused the failure to track and its correction should also be documented.
(8) Cavity Radiometers
(i) All-weather instruments: Cavity radiometers modified for continuous use should be
checked daily to ensure that all safety features are operating properly. These might
include such items as automatic shutters, rain sensors or fan switches. The manufacturer’s
operating manual should be consulted. Fans in continuous operation should be checked
for proper operation.
(ii) Fair-weather instruments: Instruments operated during fair weather conditions must
be checked for proper alignment and correct signal and power connections as part
of the set-up procedure. Shutters should be checked to ensure correct operation before
measurements begin. Following use, the exterior of the instrument should be wiped
down and the entire instrument inspected for any damage, including the lodging of
any insects within the instrument cavity. If the instrument is moved into a heated enclosure
following the measurement period, care should be taken to avoid moisture condensing
in the cavity. Although the transducer coating is not water soluble, over time, chemical
constituents within the condensing liquid can cause changes in the absorptance of
the coating. Cleaning of the sensor should be done only by qualified personnel.
(9) Shaded Instruments - Diffuse Irradiance, Infrared Irradiance
Each shaded instrument must be checked to ensure that the shading device completely covers
the outer dome of the instrument. These checks are similar to those above for the direct beam
instrum ents.
(10) Data acquisition/computer systems
The system collecting the data should be checked to ensure that it is operational. The operator,
in conjunction with the site scientist, should devise appropriate methods to decide whether
the system is operational. Simply looking at a computer screen is NOT sufficient. Tests should
be devised to detect that data are being acquired successfully, that the time stamp is correct
and that the system has not malfunctioned since the last check.
A correct system time is crucial because data are being obtained at one second intervals.
Unfortunately, many PC compatible computers have very poor clock systems. Each day the
clock offset should be recorded and the time corrected if this offset is greater than one second.
If the clock varies by more than 10 seconds per day, a new clock should be installed. A system
changing at a rate of less than one second per month would be ideal (see section 2.3.1).
(11) Where possible, the site operator should be able to review the data from the previous day.
This information will allow him/her to detect any significant changes that may have occurred
during the day. An example of such a change would be a passive tracker that was not level.
During the morning when the observer checks the shading of instruments it would be found
correct, but during the afternoon the diffuse flux would increase because of the shading disk
moving off the sensor.
Comentarios a estos manuales