Kippzonen BSRN Scientific Solar Monitoring System Manual de usuario Pagina 58

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 188
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 57
46
to a fault in the system by performing the same zero test with the resistor attached
directly to the input terminal of the unit. Servicing by authorized personnel is required
if the data acquisition unit fails. If the unit does require servicing it is also a good
opportunity to have the unit calibrated, a procedure that should be repeated every two
years. If the problem is not found in the data acquisition unit, it must be assumed that
local conditions are causing electrical interference. The cabling should be rerouted
and the test repeated. Interference can be reduced by keeping signal cables away
from power cables. It is good practice to avoid parallel routing and to intersect all cables
at 90 degrees whenever possible.
(ii) Measure the lead resistance. Short the resistor and measure the resistance of the leads
as seen by the data acquisition system (a bridge circuit may have to be built for this
test depending upon the capabilities of the system). The resistance should be less
than 10 ohms. If the resistance is satisfactory, the resistor (and bridge) can be removed
from the circuit and the pyranom eter returned. If the value is excessive, determine if
this is a result of the length of the cabling. This can be accomplished by calculating
the overall resistance of the cable by either measuring a short length of similar cable
or obtaining the specification of resistance per unit length from the manufacturer. Once
the unit length is obtained, an approximate value of the entire length can be calculated.
If the resistance can be attributed to the length of the cable one can account for the
commensurate loss in voltage when the measurement is converted into engineering
units. If the resistance is greater than indicated by the length of the cable, it is caused
by a fault within the cable. This must then be repaired or replaced.
(iii) Test the complete system. Measure the resistance of the pyranometer as installed
with the data acquisition system and check that it is approximately within the
manufacturer's specification. This has to be done at night or with the dome covered
unless the resistance measurement is in the offset compensation mode in which case
it would be unaffected by the pyranometer signal voltage.
(2) Installation of constant signals
For greater assurance in the reliability of the data, channels should be set aside to be used
with known signals in the same range as those signals being measured. For example, on
a system where both resistance and voltages are being measured, a fixed resistor and a
known voltage should be included as part of the sampling sequence. These provide a means
of rapidly flagging any changes in the quality of the measurements. Obviously, in a multiplexed
system, the potential exists for scanning problems being missed.
(3) Instrument resistance checks
An easy means of determining a fault in cabling or a sensor which outputs a voltage is testing
for changes in the resistance. On programmable systems, it is encouraged to test the resistance
of each sensor on a daily basis to determine if any significant changes have occurred in the
overall resistance of each channel. While a trained operator may observe changes in a signal
associated with a broken wire or instrument, floating channels are not necessarily easy to
detect immediately if they are not well defined (e.g., pyrgeometer voltages) or adjacent to
a channel with a similar signal.
(4) Programmable flagging
If the system is capable of automatic data quality checks, it should be programmed to set
flags when:
(i) any irradiance values fall outside the range -10 <E< 1200 Wm . The program should
-2
not be designed to delete any data autom atically.
(ii) any air temperature value (air, case, dome, etc.) is outside the normal climatological
range of the station by greater than ± 10 °C. The program should not be designed to
delete any data automatically.
Vista de pagina 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 187 188

Comentarios a estos manuales

Sin comentarios