94
Spectral selectivity ± 2%
percentage deviation of the
product of spectral absorptance
and spectral transmittance
from the corresponding mean
within 0.35 :m and 1.5 :m
Temperature response ± 1%
percentage deviation due to
change in ambient temperature
within an interval of -20 to +50 °C,
relative to 20 °C.
Tilt response 0.25%
percentage deviation from the
responsivity at 00 tilt (horizontal)
due to change in tilt from 0° to 90°
at 1000 W m irradiance
-2
Viewing angle 2 B sr
Irradiance 0 - 1400 W m (max.4000 W m )
-2 -2
Spectral range 305-2800 nm (50% points)
335-2200 nm (95% points)
Sensitivity between 7 and 25 :V W m
-1 2
Impedance 40-100 Ohm
Receiver paint Carbon black
Glass domes Schott K5 optical glass 2 mm thick, 30 mm and 50 mm outer
diameter
Desiccant Silicagel
Spirit level Sensitivity 0.1° (bubble half out of the ring) Coincide with base
of the instrument. Detector surface and base are coplanar within
0.
Materials Anodized aluminium case Stainless steel screws in stainless
steel bushes. White plastic screen of ASA
Drying cartridge PMMA
Weight 830 g
Cable length 10 m
Dimensions 91.5 mm total height, 150 mm diameter, 25 mm dome height,
50 mm dome diameter
B 2.4 Kipp and Zonen Delft BV PYRANOMETER CM 31 (additions/changes to CM 21)
Spectral Range 200-4000 nm (50% points)
290-3500 nm (95% points)
Spectral selectivity max. 2% in the spectral range 300 to 3000 nm
Zero off-set + 4 W m
-2
response to 200 W m